==== Front Ann Saudi MedAnn Saudi MedAnnals of Saudi Medicine0256-49470975-4466King Faisal Specialist Hospital and Research Centre 1686186310.5144/0256-4947.2006.216asm-3-216Brief ReportInitial experience with intracardiac echocardiography guidance in transcatheter closure of interatrial communication (atrial septal defect, patent foramen ovale, fenestrated Fontan) in Saudi Arabia Alobaidan Mashail Moamenah Tarik Abdullah Hafeez From the Paediatric Cardiology Department, Prince Sultan Cardiac Center, Riyadh, Saudi ArabiaMashail Alobaidan, MD, Paediatric Cardiology Department Prince Sultan Cardiac Center, P.O. Box 99911, Riyadh 11625, Saudi Arabia, Fax/Tel: +966-1-473 0049, mashaila@hotmail.comMay-Jun 2006 26 3 216 219 01 12 2005 Copyright © 2006, Annals of Saudi Medicine2006This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. ==== Body The recent development of intracardiac echocardiography (ICE) has added a further dimension to the imaging modalities available to the cardiologist. Intracardiac ultrasound or echocardiography has been in existence since the 1970s with clinical use for visualization of the coronary arteries since the 1980s.1,2 Early ultrasound-tipped catheters used high frequency for detailed imaging of the coronary artery lumen and wall. However, these ultrasound-tipped catheters lacked good depth penetration, and color and pulse Doppler capabilities. With the introduction of lower frequency catheters, an increase in depth penetration was possible. The initial experience with the use of the catheter for guiding device closure of secundum atrial septal defect (ASD) and patent foramen ovale (PFO) has been recently described. 3 We report our experience using this new modality in guiding trans-catheter closure of ASD, PFO and fenestrated Fontan. We avoided transesophageal echocardiography (TEE) and general anesthesia (except for that needed for comparison of TEE and ICE). We assessed the feasibility and accuracy of ICE in guiding the percutaneous closure of intra-atrial communications and the success of device deployment at Prince Sultan Cardiac Center, Riyadh, Saudi Arabia. Patients and Methods High quality ICE imaging was obtained using a 10Fr (3.2 mm), 5.5- to 10-MHz ultrasound tipped catheter (Acuson Corporation). The advantages of this catheter include the presence of a four-way tip articulation for easy maneuverability inside the cardiac chamber and Doppler capabilities. The catheter was introduced via 10Fr sheath from the right or left femoral vein. The catheter was connected to Acuson portable machine. The catheter has a quadirectional steerable tip that houses a side-firming, multiple frequency (5–10 MHz) phased-array transducer with two-dimensional and color Doppler imaging modes. The scan plane is along the axis of the catheter with a range of tissue penetration from 2 to 120 mm. On entering the RA the transducer was rotated anteriorly to obtain an image of the tricuspid valve and right ventricle (RV) inlet. The transducer tip was then flexed posteriorly and rotated clockwise to view the aorta and atrial septum in the long axis. Movement of the catheter cranially and caudally, afforded excellent views of the superior and inferior aspects of the atrial septum. The course of the right-sided pulmonary views could be imaged from the junction of the RA and superior venacava by rotating the catheter clockwise. A short-axis image of the aortic valve and atrial septum was achieved by further posterior flexing and rotating the transducer clockwise, such that the catheter tip was directed toward the TV. The amplatzer septal or PFO occluders were used in all patients. Thirty-nine patients included 13 males, 26 females, 13 PFO closure, 20 ASD II (5 no device + 15 closure), 5 fenestration closure, and 1 fenestration creation. The patient ages ranged from 4 years to 51 years with a mean of 23.5 years. Body weight ranged from 17.2 kilograms to 94 kilograms with a mean of 49 kilograms. Results The ASD cases included 17 patients, 13 of whom were totally occluded with the device (9 with ICE alone, 4 under ICE + TEE). The device used was an Amplatzer device size between 8 and 40 mm (median 24 mm). ICE was used in sizing the ASD and post-device closure. The remaining 4 patients with ASD II failed the closure and were sent for surgical repair. There were 12 patients with PFO occluded under ICE, and we used the Amplatzer device, size 35 mm and 25 mm, cribriform occluder to close the PFO in all patients with or without aneurysm. Four patients with a fenestrated Fontan were occluded under ICE and we attempted 1 fenestration creation under ICE alone using the Amplatzer ASD device occluder size 4 to 7 mm. ICE was done before to measure the size of the fenestration and after the device was implanted to see the leak. Discussion As a result of the ease of implantation and the superior success rate of ASD closure with the Amplatzer septal occluder,4,5,6 the transcatheter occlusion of ASD has become widespread and has replaced the routine surgical closure in a lot of centers. The practice of transcatheter closure of ASD is usually performed under general anaesthesia guided by continuous monitoring of TEE. To avoid the need for general anesthesia and for best assessment of the interatrial septum (IAS), ICE was thought to be superior to conventional TEE, and recent study results clearly demonstrate that ICE monitoring in transcatheter closure is superior to TEE.7,8,9 ICE monitoring needs no general anasthesia and permits unlimitted echocardiographic viewing in fully conscious and compliant patients. ICE uses a 10-MHZ tranducer which provides a much higher image resolution than TEE, a fact that also helps to effectively respond to any complication.10 There is no doubt that ICE reduces the stress to the patient and reduces the fluoroscopy time as well as the full length of the procedure. For the interventional cardiologist it has great advantages. In this study we reported on the use of ICE and assess its accuracy in evaluating the IAS and the surrounding structures compared to TEE monitoring in guiding transcatheter device closure of the intratrial defect (ASD, PFO, fenestration). Though the number of patients in which TEE was used was small, there is close agreement between TEE and ICE in their assessment of the IAS and the adequacy of the device position. ICE was found to be superior to TEE in detecting the procedural events that may affect the success of the implantation. These data favor the use of ICE to guide the routine closure of most of ASDs in adult and children without the need for TEE and general anesthesia Initially we felt that TEE was superior in evaluating the IAS, probably because of the limited experience with the use of ICE and because we are more familiar with the use and views of the TEE. However, with time and after comparing both tools, we found it is much easier to get the views and to evaluate the atrial septal rims, except for the antero-superior rim, and the septal flap. Device positioning and procedural events were easier with ICE rather than TEE and pulmonary veins view were much clearer with ICE than the TEE view. Device impingement in both the AV valve (tricuspid and mitral valve) is viewed in both ICE and TEE with almost the same quality. Our study is comparable with Mullin’s,7 Bartel’s,8 and Michael’s9 studies. Mullin et al recently reported successful closure of ASD under guidance with ICE and found that ICE is superior to TEE for evaluation of IAS, number of ASDs and device positioning. However Hijazi in 2001 reported on the failure of ICE to detect the malposition of the device. 11 Padol reported the Italian experience of using ICE and their preference for using it in an interventional procedure.12 ICE was used recently in adults to evaluate the mitral valve (MV) and guided MV balloon valvuloplasty, as well as radiofrequency ablation by an electrophysiologist.13,14,15,16 We therefore conclude that ICE is a safe tool for guiding device closure of the interatrial defect, with better patient tolerance. However, we need more practice with with ICE to facilitate interpretation. We found that ICE provides adequate views and unique images of the atrial communication defect and we could measure the size of the ASD, PFO and fenestrated Fontan accurately. Figure 1 Atrial septal defect with characteristic left-right shunt (LA=left atrium, LRS=left-to-right shunt, RA=right atrium).. Figure 2 Split-screen imaging with simultaneous flouroscopic and intracardiac echocardiographic image. Figure 3 Intracardiac echocardiographic imaging pre- and post-Amplatze deployment. Table 1 Details of diseases in relation to devices and procedures. No. of Patients Disease Device size Procedure 12 PFO 35 PFO ICE 2 ASD II 35 (cribriform) ICE 3 Fenestration closure 4 mm ICE 2 ASD II 40 mm ICE + TEE 3 ASD II 26 mm ICE + TEE 1 ASD II 20 mm ICE 2 ASD II 16 mm ICE 1 ASD II 15 mm ICE 1 ASD II 8 mm ICE 1 Fontan I 6 mm ICE 1 Fontan I 7 mm ICE 2 ASD II 25 ( cribriform) ICE 1 PFO 25 ( cribriform) ICE 1 ASD II 14 mm ICE 5 ASD II No device (failed) ICE + TEE 1 Fenestration fontan No device ICE Table 2 Atrial septal defect (ASD) assessment by intracardiac echocardiography (ICE) and transesophageal echocardiography (TEE). Patient ASD (age) AIAS Diameter (mm) Device (mm) Flouro time (min) ICE TEE Stretched M 12 yrs Absent 7 - 8 8 17 F 12 yrs Absent 24 25 24 26 15.1 F 35 yrs Absent 38 38 40 40 23.8 F 8 yrs Absent 14 13 158 16 22.4 F 6 ½ yrs Present 11 - 14 15 4.7 F 12 yrs Present 22 - 25 25(cribiform) 11.2 M 42 yrs Absent 39 40 40 40 11.1 F 39 yrs Present 31 33 34 35 7.1 F 4 yrs Absent 15 - 15 16(cribiform) 9.2 M 45 yrs Present 34 - 34 35(cribiform) 27.5 F 13 yrs Absent 23 22 26 - 8.9 F 46 yrs. Absent 42 39 42 - 2.4 F 11 yrs Present 18 20 24 - 16.5 M 13 yrs Absent 25 - 25 26 12.2 F 9 yrs Absent 22&2 - 23 & 2 25 13.4 F 10 yrs Absent 12 - 12.5 14 14.5 F 26 yrs Present 37 - 38 - M 5 yrs Absent 19 - 19 20 28.7 F 48 yrs Absent 24 - 25 26 9.8 F 27 yrs Absent 38 - 41 - * ASD II = atrial septal defect, AIAS = aneurysm interatrial septum ==== Refs References 1 Roelandt JR Serruys PW Born N Gussenhoven WG Lancee CT tenHoff H Intravascular real-time, two-dimentional echocardiography Int J Card Imaging 1989 4 63 67 2746020 2 Bartel T Eggebrecht H Ebradlidze T Images in cardiovascular medicine. Optimal guidance for intimal flap fenestration in aortic dissection by transvenous two-dimentinal and Doppler Ultrasonogrphy Circulation 2003 107 17 18 12515735 3 Koenig P Cao QL Heitscmidt M Waight DJ Hijazi ZM Role of intracardiac Echocardiographic Guidance in Transcatheter Closure of Atrial septal defects and Patent Foramen Ovale using Amplatzer device J Interven Cardiol 2003 16 51 62 4 Fischer G Stieh J Uebing A Hoffmann U Morf G Kramer HH Experience with transcatheter closure of secundum atrial septal defects using the Amplatzer septal occluder: a single centre study in 236 consecutive patients Heart 2003 89 199 204 12527678 5 Butera G De Rosa G Chessa M Rosti L Negura DG Luciane P Transcatheter closure of atrial septal defect in young children J Am Coll Cardiol 2003 42 241 5 12875758 6 Batthish M McCrindle B Stuff it or Stitch it: Innovations in the Closure of Atrial Septal Defects University of Toronto Medical J 2003 80 3 221 6 7 Mullen MJ Dias BF Walker Fiona Siu SC Benson LN Mc Laughlin PR Intracardiac echocardiography guided device closure of atrial septal defects J Am Coll Cardiol 2003 41 285 92 12535824 8 Bartel T Konorza T Arjumand J Intracardiography is superior to conventional monitoring for guiding closure of interatrial communications Circulation 2003 107 795 7 12591745 9 Earing MG Cabalka AK Seward JB Bruce CJ Intracardiac echocardiographic guidance during transcatheter device closure of atrial septal defect and patent foramen ovale Mayo Clin Proc 2004 79 24 34 14708945 10 Bartel T Konorza T Erbel R Intracardiac Echocardiography as a new Guiding Tool for Device Closure of Interatrial Communications Science Echo 2004 71 73 11 Hijazi ZM Wang Z Cao QL Transcatheter closure of atrial septal defects and patent foramen ovale under intracardic echocardiographic guidance: feasibility and comparison with trans esophageal echocardiography Catheter Cardiovasc Interv 2001 52 2 194 9 11170327 12 Ussia GP Privitera A Campisi M Carminati M De Luca F Intracardiac echocardiography using the AcuNav ultrasound catheter during percutaneous closure of multiple atrial septal defects Ital Heart J 2004 5 5 392 5 15185905 13 Green NE Hansgen AR Carroll JD Initial clinical experience with intracardiac echocardiography guiding balloon mitral valvuloplasty: Technique, safety, utility, and Limitations Catheter Cardiovasc Interv 2004 11 63 3 385 94 15505848 14 Marrouche NF Natale A Monitoring Pulmonary Vein Isolation Using Phased Array Intracardiac Echocardiography in a Patient with Atrial Fibrillation Electromedica 2002 70 1 77 80 15 Marrouche NF Martin DO Wazni O Phasedarray intracardiac echocardiography monitoring during pulmonary vein isolation in patients with atrial fibrillation: impact on outcome and complications Circulation 2003 107 2710 6 12756153 16 Daoud EG Kalbfleisch SJ Hummel JD Intracardiac echocardiography to guide transseptal left heart catheterization for radiofrequency catheter ablation J Cardiovasc Electrophysiol 1999 10 3 358 63 10210498