==== Front Ann Saudi MedAnn Saudi MedAnnals of Saudi Medicine0256-49470975-4466King Faisal Specialist Hospital and Research Centre 1805912810.5144/0256-4947.2007.415asm-6-415Original ArticleComparison of gated SPECT, echocardiography and cardiac magnetic resonance imaging for the assessment of left ventricular ejection fraction and volumes Demir Hakan *Tan Yusuf Z. *Kozdag Guliz †Isgoren Serkan *Anik Yonca ‡Ural Dilek †Demirci Ali ‡Berk Fatma * * Department of Nuclear Medicine, Kocaeli University School of Medicine, Kocaeli, Turkey † Department of Cardiology, Kocaeli University School of Medicine, Kocaeli, Turkey ‡ Department of Radiology, Kocaeli University School of Medicine, Kocaeli, TurkeyCorrespondence and reprint requests: Hakan Demir, MD, Nuclear Medicine, Kocaeli University School of Medicine, Umuttepe Yerleskesi, Kocaeli TR-41380, Turkey, hakandemir99@yahoo.comNov-Dec 2007 27 6 415 420 01 5 2007 Copyright © 2007, Annals of Saudi Medicine2007This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.BACKGROUND Left ventricular ejection fraction (LVEF), end-diastolic volume (EDV) and end-systolic volume (ESV) can be determined non-invasively by two-dimensional echocardiography (ECHO), gated single photon emission computed tomography (GSPECT) and cardiac magnetic resonance imaging (CMRI). This study was designed to analyze the concordance between LVEF, EDV and ESV values derived from ECHO, GSPECT and CMRI. METHODS ECHO, GSPECT and CMRI were performed in a group of 21 patients with suspected coronary artery disease. LVEF, EDV and ESV values were calculated. RESULTS The mean LVEF measured with GSPECT, ECHO and CMRI were 55.9±17.8%, 55.7±16.4% and 56.4±15.7%, respectively. The mean EDV measured with GSPECT, ECHO and CMRI were 109.2±42.4 mL, 127.5±42.2 mL and 91.1±38.0 mL, respectively. The mean ESV measured with GSPECT, ECHO and CMRI were 54.2±41.2 mL, 59.9±37.6 mL and 41.8±26.9 mL, respectively. The results of linear regression analysis showed very good correlation between LVEF and ESV values derived from GSPECT, ECHO and CMRI (r=0.91, r=0.92, r=0.97 for LVEF and r=0.86, r=0.91, r=0.91 for ESV, P<0.01). Good correlations were found between EDV values obtained from GSPECT, ECHO and CMRI (r=0.71, r=0.68, r=0.73, P<0.01). Agreement between these techniques in LVEF values was also good, but not in LV volumes, according to Bland-Altman plots. CONCLUSIONS This study showed good overall correlations between LVEF, EDV and ESV values derived from GSPECT, ECHO and CMRI. LVEF obtained from any of these three imaging modalities could be used interchangeably. However, care should be taken in comparing LV volumes. ==== Body Left ventricular function and volumes have major diagnostic and prognostic importance in patients with various cardiac diseases. 1 Nowadays several noninvasive techniques are available for this purpose, including 2- or 3-dimensional echocardiography (ECHO), cardiac magnetic resonance imaging (CMRI), radionuclide ventriculography (RNV) and myocardial gated SPECT (GSPECT).1–3 Since all of these techniques have some advantages or limitations none of them is considered a gold standard. Recently, GSPECT has been proposed for the determination of left ventricular ejection fraction (LVEF), end-dia-stolic volume (EDV) and end-systolic volume (ESV). GSPECT offers the potential advantage of combining information on myocardial function and perfusion without any extra cost, discomfort and radiation risk to patient.4–7 According to previously published studies, GSPECT correlates well with both ECHO and CMRI with respect to LVEF, LV volumes and wall motion.1,4–19 The aim of this prospective study was to analyze the concordance between LVEF, EDV and ESV values derived from GSPECT, ECHO, and CMRI in the same group of patients. METHODS Twenty-one patients with known or suspected coronary artery disease, referred for routine GSPECT imaging, were included in the study. All patients underwent ECHO and CMRI. All investigations were completed within 4 weeks. There were no cardiac events between the three studies. Cardiac medications were not changed during the study. Patients who had a history of severe valvular disease, morbid obesity, pregnancy or who had myocardial infarction, angioplasty or bypass surgery during the 6 weeks before participation in the study were excluded. GSPECT acquisition was started after 45 minutes of the intravenous injection of 555 MBq of 99mTc-methoxyisobutyl-isonitrile (99mTc-MIBI) at resting condition. Data acquisition was performed with a single-head SPECT system (ADAC Laboratories, Milpitas, California, USA) equipped with a low-energy, high-resolution collimator. A 20% window around the 140 keV energy peak of 99mTc-MIBI was used. The acquisition matrix size was 64×64×16. A total of 64 projections (step-and-shoot mode, 30 s per projection) were obtained over a 180° circular orbit. Acquisitions were gated for eight frames per cardiac cycle. The images were reconstructed using filtered back-projection. The resulting transaxial slices were re-oriented perpendicular to the heart’s long axis, yielding long- and short-axis tomograms. The LVEF and left ventricular volumes were calculated using previously validated and commercially available automated software (Auto SPECT, Autoquant, ADAC Laboratories, Milpitas, California, USA; developed at Cedars-Sinai Medical Center, Los Angeles, California, USA) from the GSPECT images. ECHO was performed with a Toshiba SSA-390-A ultrasound machine (Toshiba Medical Systems, Nasu, Japan) by one experienced cardiologist blind to the results of the GSPECT study. All images were obtained with a second harmonic mode. Two-dimensionally guided M-mode recordings were taken from the parasternal long-axis view according to the criteria of the American Society of Echocardiography. EDV, ESV and LVEF were derived from apical four-chamber and two-chamber views with the previously validated modified Simpson’s biplane discs method. 20,21 All CMRI studies were performed on a 1.5-T MR scanner (Philips Gyroscan Intera Master, Einthoven, Netherlands) with a 30 mT/m maximum gradient strength and a 150 mT/m per millisecond slew rate using a synergy body coil. Our routine CMRI protocol started with the reference images of Balanced Turbo Field Echo (B-TFE) in sagittal, axial and coronal planes, and was followed by coil sensitivity (SENSE) reference images. ECG gated, breath hold of T2 weighted B-FFE sequence (TR/TE=3.1/1.5, matrix: 256×256) in four chambers, left ventricle short axis and long axis planes were performed. Values of LVEF, EDV and ESV derived from all three investigations were compared. Continuous data were expressed as the mean±SD. The agreement between LVEF, EDV and ESV derived from GSPECT, ECHO and CMRI data was determined with linear regression (Pearson’s correlation coefficient) and Bland-Altman analysis.22 The differences between GSPECT, ECHO and CMRI measurements of LVEF, EDV, and ESV were tested for significance using a t-test for paired samples and P<0.05 was considered significant. For linear regression analysis P<0.01 was considered significant. Our University Hospital’s Scientific and Ethics Committee for Human Clinical Research approved this protocol, and all patients provided informed consent prior to enrollment into the study. RESULTS Twenty-one patients (15 male and 6 female, mean age 55±9 years, range 38–74 years) were included in the study. Fifteen patients were hypertensive. One had diabetes mellitus. Five patients had a history of myocardial infarction. No patient was excluded from the study because of the poor image quality of any imaging modality. Comparison of LVEF Results The LVEF measured with GSPECT was in the range 23% to 80%, with a mean of 55.9±17.8%. The corresponding values in ECHO were similar: the LVEF was in the range 23% to 77%, with a mean of 55.7±16.4% (Table 1). The linear regression analysis showed very good correlation (r=0.91, P=0.000) between LVEF assessed with ECHO and with GSPECT. The mean difference in LVEF measured with ECHO and GSPECT was 0.2±7%, with limits of agreement of − 13.8% to 14.4% (P=0.883, 95% confidence interval, − 3.6 to 3.0), as shown by Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. No systematic under- or overestimation of LVEF was found with GSPECT as compared with ECHO. The LVEF measured with CMRI was in the range 23% to 74%, with a mean of 56.4±15.7% (Table 1). The linear regression analysis showed very good correlation (r=0.97, P=0.000) between LVEF assessed with ECHO and with CMRI. The mean difference in LVEF measured with ECHO and CMRI was −0.74±4.1%, with limits of agreement of −8.9 to 7.4% (P=0.412, 95% confidence interval, −2.6 to 1.1), as shown by Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. There was no systematic under- or overestimation of LVEF with ECHO compared with CMRI. There was also very good correlation (r=0.92, P=0.0001) between LVEF assessed with GSPECT and with CMRI. The mean difference in LVEF measured with GSPECT and CMRI was −0.5±7%, with limits of agreement of −14.7 to 13.7% (P=0.749, 95% confidence interval, −3.7 to 2.7), as shown by Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. Also no systematic under- or overestimation of LVEF was found with GSPECT compared with CMRI. In none of the patients was LVEF measured as definitely abnormal (<40%) with one method and as definitely normal (>50%) with other method. Comparison of EDV Results The EDV measured with GSPECT was in the range 46 to 198 mL with a mean of 109.2±42.4 mL. The corresponding values in ECHO were similar: the EDV was in the range 51.9–249 mL with a mean of 127.5±42.2 mL (Table 1). The linear regression analysis showed good correlation (r=0.71, P=0.000) between EDV assessed with ECHO and with GSPECT. The mean difference in EDV measured with ECHO and GSPECT was 18.3±33.4 mL, with limits of agreement of −48.6 to 85.1 mL (P=0.021, 95% confidence interval, 3.0 to 33.5), as shown by means of Bland-Altman analysis. Despite the limits of confidence containing 100% (21/21 samples) of the data points and good correlation between GSPECT and ECHO, EDV was slightly overestimated by ECHO compared with GSPECT. The EDV measured with CMRI was in the range 18.4–151 mL with a mean of 91.1±38.0 mL (Table 1). The linear regression analysis showed good correlation (r=0.73, P=0.000) between EDV assessed with ECHO and CMRI. The mean difference in EDV measured with ECHO and CMRI was 36.4±31.5 mL, with limits of agreement of −26.6 to 99.4 mL (P=0.000, 95% confidence interval, 22.1 to 50.7), as shown by Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. The EDV was overestimated by ECHO compared with CMRI. Also there was good correlation (r=0.68, P=0.001) between EDV assessed with GSPECT and with CMRI. The mean difference in EDV measured with GSPECT and CMRI was 18.1±32 mL, with limits of agreement of −46.5 to 82.1 mL (P=0.018, 95% confidence interval, 3.4 to 32.8), as shown by means of Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. The EDV was slightly, overestimated by GSPECT compared with CMRI. Comparison of ESV Results The ESV measured with GSPECT was in the range 9 to 149 mL with a mean of 54.2±41.2 mL. The corresponding values in ECHO were similar: the ESV was in the range 14.8–164 mL with a mean of 59.9±37.6 mL (Table 1). The results of linear regression analysis showed a very good correlation (r=0.86, P=0.000) between ESV assessed with ECHO and with GSPECT. The mean difference in ESV measured with ECHO and GSPECT was 5.7±20.8 mL with limits of agreement of −36.0 to 47.3 mL (P=0.228, 95% confidence interval, −3.8 to 15.1), as shown by Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. The ESV was not under- or overestimated by GSPECT compared with ECHO. The ESV measured with CMRI was in the range 6.2 to 98 mL with a mean of 41.8±26.9 mL (Table 1). The results of linear regression analysis showed a very good correlation (r=0.91, P=0.000) between ESV assessed with ECHO and with CMRI. The mean difference in ESV measured with ECHO and CMRI was 18.1±17.3 mL with limits of agreement of −16.5 to 52.7 mL (P=0.000, 95% confidence interval, 10.2 to 25.9), as shown by means of Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data point. The ESV was underestimated by CMRI compared with ECHO. Also there was a very good correlation (r=0.91, P=0.000) between ESV assessed with GSPECT and with CMRI. The mean difference in ESV measured with GSPECT and CMRI was 12.4±19.9 mL, with limits of agreement of −27.4 to 52.2 mL (P=0.010, 95% confidence interval, 3.3 to 21.5), as shown by means of Bland-Altman analysis. These limits of confidence contained 95.2% (20/21 samples) of the data points. The ESV was underestimated by CMRI compared with GSPECT. DISCUSSION Left ventricular function and volumes have major diagnostic and prognostic importance in patients with various cardiac diseases.1 Nowadays several noninvasive techniques are available for this purpose, including ECHO, CMRI, RNV and GSPECT. In recent years all of these imaging techniques have become more available, and clinicians could be faced with LVEF and LV volume results obtained with various imaging techniques during a long follow-up period. Therefore, cardiologists need to know the exact correlation and agreement between these techniques. Another concern is whether measurements of LV volume and ejection fraction derived from different techniques could be used interchangeably. In contrast to most studies, which included only two imaging modalities (GSPECT vs. ECHO or CMRI; ECHO vs. GSPECT or CMRI) we performed three imaging methods in the same group of patients. The present study is, to our knowledge, only the second one to compare left ventricular functions (LVEF, EDV and ESV) derived from GSPECT, CMRI and ECHO in the same group of patients. In the first study, Yamamuro et al23 performed multi-detector row CT and magnetic resonance imaging in 50 patients, two-dimensional ECHO in 41 patients and GSPECT in 27 patients. They found that the standard deviation of the EF difference between multi-detector row CT and MR imaging was significantly less than that between echocardiography and MR imaging (P<0.001) or that between GSPECT and MR imaging (P<0.001). According to previous studies that compared GSPECT and ECHO there was very good agreement between LVEF, EDV and ESV values.1,5–7,9–11,23 Fleming found a high correlation (r=0.76) between resting echocardiographic EFs and SPECT resting gated sestamibi images in patients with single-vessel disease, and a moderate correlation (r=0.68 and r=0.68) in patients with 2- and 3-vessel disease, respectively.5 In a recently published study, we compared left ventricular function and volumes of GSPECT and ECHO in patients with dilated cardiomyopathy. We found good correlations for the assessment of LVEF, EDV and ESV (r=0.72, r=0.71 and r=0.71, respectively).9 In the present study, when we compared the results of GSPECT and ECHO, the values for LVEF, EDV and ESV correlated strongly (r=0.91, r=0.81, r=0.71, respectively). Also no systematic under- or overestimation of LVEF and ESV values was found with GSPECT as compared with ECHO according to Bland-Altman plots. However, EDV was slightly overestimated by ECHO compared with GSPECT. A number of studies have compared GSPECT and CMRI.4,8,12–19 Thorley et al. compared GSPECT with CMRI measurements of LVEF and EDV in 50 patients and found a good correlation (r=0.82 and r=0.90). 13 In another study, Bavelaar-Croon and colleagues also found a good correlation for LVEF, EDV and ESV (r=0.85, r=0.94, r=0.95, respectively).14 In our study, LVEF, EDV and ESV results of GSPECT were also correlated very well with CMRI results (r=0.92, r=0.68, r=0.91, respectively). LVEF was not under-or overestimated by GSPECT compared with CMRI. But, EDV and ESV were slightly underestimated by CMRI compared with GSPECT. Interestingly, most previous studies have reported that volumes were underestimated by GSPECT compared with CMRI.16,17,24 However, some studies showed results similar to ours.8,15,18 Faber et al reported that both Quantitative Gated SPECT (QGS) and Emory Cardiac Toolbox (ECTb) had given EDV and ESV 20% to 30% larger than the MR measurements. 8 Similarly, Vaduganathan et al found volumes to be larger when computed with GSPECT than with CMRI.15 Because of high spatial resolution, excluding papillary muscles and trabeculations from volumes in CMRI was easier than in either GSPECT and ECHO. In our study, to exclude major vascular structures and valves from ventricular volume slices of CMRI were minimally repositioned to midventricular region at the base of the heart. So, inclusion or exclusion of the most basal slice, which consists of parts of LV myocardium, outflow tract, and left atrium, could be the main reason for the difference between previous studies and the present study. In addition, the frame numbers per cardiac cycle were different in CMRI and GSPECT (16 vs. 8 frames for CMRI and GSPECT, respectively). We used 8 frames for GSPECT images to obtain sufficient count statistics. This could be the technical reason for the conflict in volume results. Direct comparison of CMRI and ECHO was made using different protocols and imaging techniques, in some previous studies.3,23,24–28 Hoffmann et al, in a multi-centre study, assessed the agreement of LVEF derived from cineventriculography, cardiac magnetic resonance imaging, unenhanced and contrast-enhanced echocardiography in 120 patients. They reported that the correlation for EF between MRI and unenhanced echocardiography was 0.60.3 In most of those studies that compared CMRI and ECHO, measurements of CMRI were larger than ECHO results.3,25–28 In the present study, LVEF, EDV and ESV values from CMRI and ECHO correlated very well. While LVEF was not under-or overestimated by CMRI compared with ECHO, the values of EDV and ESV were underestimated. We think the reasons for the difference between CMRI and GSPECT results (exclusion of basal portions of left ventricle and high spatial resolution of CMRI) also account for the difference between ECHO and CMRI results. In addition to very good correlations and agreements between the LVEF values for these three imaging modalities, in none of the patients was LVEF measured as definitely abnormal (<40%) with one method and as definitely normal (>50%) with the other method. This result is in agreement with the study by Bavelaar-Croon and colleagues.14 Thus, LVEF results obtained from any of these three imaging modalities could be used interchangeably. In conclusion, we found overall very good correlations and agreements between LVEF derived from GSPECT, CMRI and ECHO as shown both linear regression and Bland-Altman analysis. Even though we found good correlations between EDV and ESV values derived from all of these imaging modalities, agreemens between these results were not as good as between LVEF results. Therefore, care should be taken in comparing LV volumes derived from these three imaging methods. Table 1 Measurement of left ventricular volumes and ejection fraction with the three imaging methods. Mean LVEF (%±SD) n:21 Mean EDV (ml±SD) n:21 Mean ESV (ml±SD) n:21 GSPECT 55.9±17.8 (23–80) 109.2±42.4 (46–198) 54.2±41.2 (9–149) ECHO 55.7±16.4 (23–77) 127.5±42.2 (51.9–249) 59.9±37.6 (14.8–164) CMRI 56.4±15.7 (23–74) 91.1±38.0 (18.4–151) 41.8±26.9 (6.2–98) ==== Refs REFERENCES 1 Cwajg E Cwajg J He ZX Hwang WS Keng F Nagueh SF Gated myocardial perfusion tomography for the assessment of left ventricular function and volumes: comparison with echocardiography J Nucl Med 1999 40 1857 65 10565781 2 Niemeyer MG van der Wall EE Kuijper AF Cleophas AT Pauwels EK Nuclear cardiology, current applications in clinical practice Angiology 1995 46 591 602 7618762 3 Hoffmann R von Bardeleben S ten Cate F Borges AC Kasprzak J Firschke C Lafitte S Assessment of systolic left ventricular function: a multi-centre comparison of cineventriculography, cardiac magnetic resonance imaging, unenhanced and contrast-enhanced echocardiography Eur Heart J 2005 6 607 16 4 Stollfuss JC Haas F Matsunari I Neverve J Nekolla S Schneider-Eicke J Regional myocardial wall thickening and global ejection fraction in patients with low angiographic left ventricular ejection fraction assessed by visual and quantitative resting ECG-gated 99mTc-tetrofosmin single-photon emission tomography and magnetic resonance imaging Eur J Nucl Med 1998 25 522 30 9575249 5 Fleming RM A tete-a-tete comparison of ejection fraction and regional wall motion abnormalities as measured by echocardiography and gated sestamibi SPECT Angiology 2002 53 313 21 12025919 6 Nichols K Lefkowitz D Faber T Folks R Cooke D Garcia EV Echocardiographic validation of gated SPECT ventricular function measurements J Nucl Med 2000 41 1308 14 10945519 7 Choragudi NL Prakash AM Sun Y Prasad P Chiaramida SA Lucariello RJ Comparison of echocardiography with technetium 99m-gated single photon emission computed tomography as diagnostic tools for left ventricular ejection fraction Echocardiography 2001 18 627 32 11801203 8 Faber TL Vansant JP Pettigrew RI Galt JR Blais M Chatzimavroudis G Cooke CD Evaluation of left ventricular endocardial volumes and ejection fractions computed from gated perfusion SPECT with magnetic resonance imaging: comparison of two methods J Nucl Cardiol 2001 8 645 51 11725260 9 Berk F Isgoren S Demir H Kozdag G Sahin T Ural D Assessment of left ventricular function and volumes for patients with dilated cardiomyopathy using gated myocardial perfusion SPECT and comparison with echocardiography Nucl Med Commun 2005 26 701 10 16000988 10 Nakae I Matsuo S Koh T Mitsunami K Horie M Left ventricular systolic/diastolic function evaluated by quantitative ECG-gated SPECT: comparison with echocardiography and plasma BNP analysis Ann Nucl Med 2005 19 447 54 16248380 11 Chua T Yin LC Thiang TH Choo TB Ping DZ Leng LY Accuracy of the automated assessment of left ventricular function with gated perfusion SPECT in the presence of perfusion defects and left ventricular dysfunction: correlation with equilibrium radionuclide ventriculography and echocardiography J Nucl Cardiol 2000 7 301 11 10958271 12 Ioannidis JP Trikalinos TA Danias PG Electrocardiogram-gated single-photon emission computed tomography versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction: a meta-analysis J Am Coll Cardiol 2002 39 2059 68 12084609 13 Thorley PJ Plein S Bloomer TN Ridgway JP Sivananthan UM Comparison of 99mTc tetrofosmin gated SPECT measurements of left ventricular volumes and ejection fraction with MRI over a wide range of values Nucl Med Commun 2003 24 763 9 12813194 14 Bavelaar-Croon CD Kayser HW van der Wall EE de Roos A Dibbets-Schneider P Pauwels EK Left ventricular function: correlation of quantitative gated SPECT and MR imaging over a wide range of values Radiology 2000 217 572 5 11058662 15 Vaduganathan P He ZX Vick GW 3rd Mahmarian JJ Verani MS Evaluation of left ventricular wall motion, volumes, and ejection fraction by gated myocardial tomography with technetium 99m-labeled tetrofosmin: a comparison with cine magnetic resonance imaging J Nucl Cardiol 1999 6 3 10 10070835 16 Chan J Jenkins C Khafagi F Du L Marwick TH What is the optimal clinical technique for measurement of left ventricular volume after myocardial infarction? A comparative study of 3-dimensional echocardiography, single photon emission computed tomography, and cardiac magnetic resonance imaging J Am Soc Echocardiogr 2006 19 192 201 16455424 17 Bax JJ Lamb H Dibbets P Pelikan H Boersma E Viergever EP Comparison of gated single-photon emission computed tomography with magnetic resonance imaging for evaluation of left ventricular function in ischemic cardiomyopathy Am J Cardiol 2000 86 1299 305 11113402 18 Mochizuki T Murase K Tanaka H Kondoh T Hamamoto K Tauxe WN Assessment of left ventricular volume using ECG-gated SPECT with technetium-99m-MIBI and technetium-99m-tetrofosmin J Nucl Med 1997 38 53 7 8998150 19 Tadamura E Kudoh T Motooka M Inubushi M Okada T Kubo S Use of technetium-99m sestamibi ECG-gated single-photon emission tomography for the evaluation of left ventricular function following coronary artery bypass graft: comparison with three-dimensional magnetic resonance imaging Eur J Nucl Med 1999 26 705 12 10398818 20 Schiller NB Shah PM Crawford M DeMaria A Devereux R Feigenbaum H Recommendations for quantitation of the left ventricle by two dimensional echocardiography: American Society of Echocardiography Committee on Standards Subcommittee on Quantitation of Two-Dimensional Echocardiography J Am Soc Echocardiogr 1989 2 358 67 2698218 21 Otterstad JE Froeland G St John Sutton M Holme I Accuracy and reproducibility of biplane two-dimensional echocardiographic measurements of left ventricular dimensions and function Eur Heart J 1997 18 507 13 9076390 22 Bland JM Altman DG Statistical methods for assessing agreement between two methods of clinical assessment Lancet 1986 1 307 10 2868172 23 Yamamuro M Tadamura E Kubo S Toyoda H Nishina T Ohba M Cardiac functional analysis with multi-detector row CT and segmental reconstruction algorithm: comparison with echocardiography, SPECT, and MR imaging Radiology 2005 234 381 90 15670995 24 Tadamura E Kudoh T Motooka M Inubushi M Shirakawa S Hattori N Assessment of regional and global left ventricular function by reinjection T1-201 and rest Tc-99m sestamibi ECG-gated SPECT: comparison with three-dimensional magnetic resonance imaging J Am Coll Cardiol 1999 33 991 97 10091826 25 Gutierrez-Chico JL Zamorano JL Perez de Isla L Orejas M Almeria C Rodrigo JL Comparison of left ventricular volumes and ejection fractions measured by three-dimensional echocardiography versus by two-dimensional echocardiography and cardiac magnetic resonance in patients with various cardiomyopathies Am J Cardiol 2005 95 809 13 15757621 26 Malm S Frigstad S Sagberg E Larsson H Skjaerpe T Accurate and reproducible measurement of left ventricular volume and ejection fraction by contrast echocardiography: a comparison with magnetic resonance imaging J Am Coll Cardiol 2004 44 1030 5 15337215 27 Heuschmid M Rothfuss JK Schroeder S Fenchel M Stauder N Burgstahler C Assessment of left ventricular myocardial function using 16-slice multidetector-row computed tomography: comparison with magnetic resonance imaging and echocardiography Eur Radiol 2006 16 551 9 16215736 28 Jenkins C Bricknell K Hanekom L Marwick TH Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography J Am Coll Cardiol 2004 44 878 86 15312875